Respiratory Therapist

Category - Infection Control and Equipment Troubleshooting and Quality Control

Select all of the following circumstances that could affect the accuracy of a galvanic fuel cell that is placed inline a ventilator circuit.
  1. Excess water on the cell membrane
  2. Altitude
  3. Temperature of the circuit
  4. Excessive pressure settings.
Explanation
Answers: A, B, and D - An inline galvanic fuel cell placed inline a ventilator circuit can be affected by excess water on the cell membrane, altitude, or excessive pressure settings.

Excess water in the circuits can cause excess water on the cell membrane, which will affect accuracy by interfering with the oxygen diffusing, and a low reading will occur. Altitude can also affect the galvanic fuel cell as it is affected by barometric pressure. Excessive pressure will cause the device to have a higher than accurate reading.
Was this helpful? Upvote!
Login to contribute your own answer or details